5.2 Temperature Controller Unit (TCU)
5.2.1. TCU startup
1. make sure the laser power supply is disconnected (see section 3.3 for connec-
tions)
2. turn on the TCU
3. set the desired temperature:
a) select Setting oC using the switch (3) shown in figure 5.2 to display the
desired temperature
b) turn the knob (5) to set the temperature to the desired value
c) select Real oC using the switch (3) to display the actual temperature
4. allow the laser temperature to reach the temperature setting
CAUTION:
Do not operate the laser at a temperature above the values specified
in the ALPES LASERS datasheet. Permanent damage may result.
5. connect back the laser power supply
5.2.2. Internal/external mode
In internal mode, the temperature setting is performed using the temperature knob
of the TCU front panel. In external mode, the temperature setting, monitoring and
interlock are performed by an external control system using the BNC connections of
the rear panel. In that case:
1. set the switch (5) to external
2. provide a ±6.5V power supply to the External Ref BNC connector, using 100
mV/
C
3. read out the temperature through the Monitoring BNC connector, providing 10
mV/
C.
c
2014 SA 65