Chapter 7: Maintenance
6. Oscillator check (refer to figure 7.13): measure the frequency on the pin 7 of
U13 (1); the value must be 28KHz + /- 3KHz.
7. Current/Voltage ratio adjustment (refer to figure 7.13)
a) Adjust the trimmer P4 (6) in order to read B4 = 0.6V (4).
b) Set the selector (5) to mode Setting +I: the LCD screen should display
1.00 A.
Note: If it is not the case, adjust it with P2 (2).
c) Adjust the trimmer P4 in order to read B4 = 3.0 V: the LCD screen should
display 5.00 A.
Note: If it is not the case, adjust it with P2 (2). Example : 0,6 V 0,99 A
and 3V 4,99 A.
d) Set the switch on the front panel to mode Setting -I.
e) Repeat steps [2] and [4] for P5 (7) and B5 (3).
1
2
3
4
Figure 7.13.: TCU main board, oscillator check (left), current/voltage ratio adjustment.
(right).
96 QCL user’s manual v3.1.3