Alpes Lasers . TC-3 Datasheet Page 71

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 70
5.1 Temperature Controller TC-3
Once you have made all your adjustments, press the MENU button to exit the menu
(in the future, if you are only changing one value, you do not need to press the knob
— pressing the MENU button while changing a value will save the value and exit).
Next, connect the cable between your housing and the Output connector of the TC-
3. We recommend using our cables as they have been designed to work well with the
TC-3. If using your own cables, ensure they have been properly wired according to
the pin-out of the TECSource and your fixture.
Set the set point to a temperature close to room temperature (typically +15
C) and
press the Output button. The output will turn on and you will see the voltage and
current begin driving the fixture to the set point you have chosen. Depending on the
thermal size of your fixture, it may take seconds or several minutes to reach the set
point.
If you notice the temperature is oscillating around the set point and not stabilizing,
you may need to adjust the Gain setting in the menu. If the temperature is quickly
jumping up and down, the Gain will typically need to be reduced. If the temperature
is slowly moving up and down, try a higher Gain. You may need to experiment with
several gain settings to find the ideal value, and for even finer control, you can set the
Gain to PID and directly set the PID control values.
Finally, adjust the set point downwards or upwards to the desired set point. It is
good practice to adjust the set point in steps of 5
C and wait for the actual tempera-
ture to stabilize before changing the set point again, as the HHL and TO3 housing age
prematurely with large temperature swings. The LLH is less sensitive and can with-
stand large temperature swings. The recommended temperature change rates are
given in table 5.3. Using the settings of table 5.2 ensures these rates are respected.
CAUTION:
The TC-3 contains a feature known as AutoTune. Usage of this feature
is NOT RECOMMENDED with the housings of Alpes Lasers. Using
AutoTune risks damaging or reducing the lifetime of the Peltier
elements.
Housing I
lim
HHL (single stage) 3.0 A
HHL (double stage) 2.5 A
LLH 5.0 A
Table 5.1.: Current limits settings
c