Alpes Lasers . TC-3 Datasheet Page 50

  • Download
  • Add to my manuals
  • Print
  • Page
    / 126
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 49
Chapter 3: Installation
3.3.7. TC-LLH cable
The TC-3 controls the temperature of the LLH with a Peltier junction and monitors
the temperature with a PT100 through the TC-LLH cable. The Peltier connector is
identical to the connector shown in the left side of Fig. 3.10. The controller connector
is a DB-15 connector described in Table 3.6. The Fans, ID and Remote Sensor pins
are not used by the ALPES LASERS products.
Pin number Description
1, 2 & 9 TE (+)
3, 4 & 10 TE (-)
5 & 6 Ground
7 Sensor +
8 Sensor -
11 Fan +
12 Fan -
13 ID
14 Remote Sensor+
15 Remote Sensor -
Table 3.6.: TC-LLH cable connector pinout.
3.3.8. TC-HHL cable
The TC-3 controls the temperature of the HHL with a Peltier junction and monitors the
temperature with a PT100 or NTC sensor through the TC-HHL cable. The Peltier con-
nector is identical to the connector shown in the left side of Fig. 3.10. The controller
connector is a DB-15 connector described in Table 3.6. The Fans, ID and Remote
Sensor pins are not used by the ALPES LASERS products.
3.3.9. CPL cable
The CPL cable shown in figure 3.11 connects the LDD to the TPG and external power
supply, using a DE-9 connector on the LDD side, yellow/black banana connectors for
Bias-T operation, Lemo 00 for LDD power supply, BNC for LDD timing signal and
red/black banana connectors for TPG power supply
The LDD side consists of a D-sub-9 connector, composed of 9 pins as shown in
figure 3.12 and described in table 3.7.
40 QCL user’s manual v3.1.3